Roche Cobas c311


The cobas c 311 analyzer is a stand alone system that offers consolidated testing from a broad menu of clinical chemistry applications. This analyzer has the capacity for ion-selective electrode (ISE) determination of sodium, potassium, and chloride in serum, plasma, and urine. In addition, measurement of HbA1c levels in whole blood can also be performed, making it a truly flexible analyzer.
Copy Files
- There are 3 files need to be copied and replace.
- License File
- SMC Files
- Plugin Files
- Firstly, stop all these services.
- Server Console (only when updating sys-services-config.properties file)
- Paste the created license into 5 different folder
- My Computer > C: > Chemolims > License
- My Computer > C: > Chemolims > server > minimal > license
- My Computer > C: > Chemolims > console > license
- My Computer > C: > Windows > Syswow64 > license
- My Computer > C: > Windows > System32 > license
- Go to My Computer > C: > Chemolims > License and double click on update_license.

- Next, open Chemolims > console > bin > plugins

- Replace .jar files for the old connector
- Paste .jar file for the new connector
- Copy the existing .config and .log4j files inside C:/ChemoLIMS/console/bin/plugins
- Then, right click on bio-rad-d10-config and select edit with notepad ++.
- Make sure to change from bio-rad-d10 to sysmex-xn1000 and then Save.

- Then, right click on bio-rad-d10-logj4 and select edit with notepad++.
- Make sure to change from bio-rad-d10 to sysmex-xn1000 and then Save.

- Then, right click on recreate_executable and select edit with notepad++

- Add the new connector sysmex-xn1000-plugin .exe file name and save. (Remark: Analyzer name should be exactly the same with the name inside the excel file)

- Then double click on the recreate_executable.
- There should have 4 files for instrument
- .exe
- .jar
- .log4j
- config
- Then copy 8 SMC files from Google Drive and paste inside C: > Chemolims > console > bin

- Then right-click on sys-services-config.properties and select edit with notepad ++

- Add Symex XN-1000 and then save. (Remark: Analyzer name should be exactly the same with the name inside the excel file)

- Lastly, double click the recreate_executeble inside C: > Chemolims > console > bin
Mapping
- Open pgAdmin III

- Once open pgAdmin, double click on PostgreSQL 8.4 as below

- Then double click on the Databases as below

- Then double click on chemolims as below

- Then double click on Schemas as below

- Then double click on public as below

- Then double click on Tables as below

- Then find a table name analyzer_masters. Right-click and select View Data > View All Row

- Check either the instrument for Mission U500 is already in the table or not.
- If Yes, can proceed to the next step.
- If No, add all details at the * row and click save.

- Then find a table name analyzer_test. Right-click and select View Data > View All Row

- Check either the test for Mission U500 is already in the table or not.
- If Yes, can proceed to the next step.
- If No, Then find a table name test_masters. Right-click and select View Data > View All Row

- Scroll down and find the test name from the list and copy the details and paste in the analyzer_test table.

- Please make sure to paste at the correct column and the analyzerid column must be similar to the analyzer_masters table.

- For analcmdcode and analtestcode column, need to get the details from the machine itself.

Configuration
- First go to Menu List > Utility > System
- Change the setting to be exactly the same as in the picture.

Test Database
- Start ChemoLIMS Server Console

- Wait until the Connector Status change to Running.
(Remarks: Update test database)