Install and Configure Bio-Rad D-10

Install and Configure Bio-Rad D-10

D-10 Hemoglobin Analyzer

The D-10™ instrument provides comprehensive, fully automated hemoglobin testing in a compact footprint. Combining diabetes and β-thalassemia testing on a single platform, the system offers simple, efficient operation for clinics, physician offices or clinical laboratories.

Copy Files

  1. There are 3 files need to be copied and replace.
    • License File
    • SMC Files
    • Plugin Files
  2. Firstly, stop all these services.
    • Server Console (only when updating sys-services-config.properties file)
  3. Paste the created license into 5 different folder
    • My Computer > C: > Chemolims > License
    • My Computer > C: > Chemolims > server > minimal > license
    • My Computer > C: > Chemolims > console > license
    • My Computer > C: > Windows > Syswow64 > license
    • My Computer > C: > Windows > System32 > license
  4. Go to My Computer > C: > Chemolims > License and double click on update_license.
  5. Next, open Chemolims > console > bin > plugins
  6. Replace .jar files for the old connector
  7. Paste .jar file for the new connector
  8. Copy the existing .config and .log4j files inside C:/ChemoLIMS/console/bin/plugins
  9. Then, right click on bio-rad-d10-config and select edit with notepad ++.
  10. Then make sure the highlighted name change according to analyzer name and Save.
  11. Then, right click on bio-rad-d10-logj4 and select edit with notepad++.
  12. Make sure the analyzer is correct and then Save.
  13. Then, right click on recreate_executable and select edit with notepad++
  14. Add the new connector .exe file name and save. For example biorad-d10-plugin.exe (Remark: Analyzer name should be exactly the same with the name inside the excel file)
  15. Then double click on the recreate_executable.
  16. There should have 4 files for instrument
    • .exe
    • .jar
    • .log4j
    • config
  17. Then copy 8 SMC files from Google Drive and paste inside C: > Chemolims > console > bin
  18. Then right-click on sys-services-config.properties and select edit with notepad ++
  19. Add the new connection name and then save. For example Bio-Rad D-10 (Remark: Analyzer name should be exactly the same with the name inside the excel file)
  20. Lastly, double click the recreate_executeble inside  C: > Chemolims > console > bin

Mapping

  1. Open pgAdmin III 
  2. Once open pgAdmin, double click on PostgreSQL 8.4 as below
  3. Then double click on the Databases as below
  4. Then double click on chemolims as below
  5. Then double click on Schemas as below
  6. Then double click on public as below
  7. Then double click on Tables as below
  8. Then find a table name analyzer_masters. Right-click and select View Data > View All Row
  9. Check either the instrument for Bio-rad D10 is already in the table or not.
    1. If Yes, can proceed to the next step.
    2. If No, add all details at the * row and click save.
  10. Then find a table name analyzer_test. Right-click and select View Data > View All Row
  11. Check either the test for Bio-rad D10 is already in the table or not.
    1. If Yes, can proceed to the next step.
    2. If No, Then find a table name test_masters. Right-click and select View Data > View All Row
    3. Scroll down and find the test name from the list and copy the details and paste in the analyzer_test table.
    4. Please make sure to paste at the correct column and the analyzerid column must be similar to the analyzer_masters table.
    5. For analcmdcode and analtestcode column, need to get the details from the machine itself.

Connectivity

  1. Make sure the device in the Sleep Mode.
  2. Next, select the Setting tab and select Next to go page 3.
  3. Select Export to LIS and make sure it is in YES and tick on After each injection.

Test Database

  1. Start ChemoLIMS Server Console
  2. Wait until the Connector Status change to Running.
  3. Select Data Tab and highlight the Sample ID that wants to be export into ChemoLIMS and select Export.